Z-2714-2026: OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

Recall record Z-2714-2026 for Integra LifeSciences Corp. (NeuroSciences). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2714-2026
Company in FDA record
Integra LifeSciences Corp. (NeuroSciences)
FDA classification date
Jul 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

Source facts

Action
Integra issued an URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice to its consignees on 06/05/2026 via overnight mail. The notice explained the issue, hazards and risks, and requested the following: "Customer/Medical Facility-Actions to Take: 1. If your facility has validated its own sterilization process, it is in your discretion to follow your validated process recognizing that Integra LifeSciences has only validated the sterilization process noted below. 2. If utilizing the Gravity Sterilization Method, stop using this method immediately. The Gravity Sterilization Method for sterilizing Omni-Tract devices is not effective. 3. Replace the instructions in the current IFUs (40-9015 Rev BAA and 40-6008 Rev AAA) for 'sterilization' sections with the following: Sterilization Following proper cleaning and prior to sterilization, use an appropriate sterilization wrap with sterilization tray to ensure that components and devices are not damaged in transportation to and from the sterilizer. NOTE: Validation was completed using a wrapped sterilization tray and not a rigid container. Steam Sterilization Method: Process Validation procedures followed ANSI/AAMI ST79 standards for Prevacuum. Steam Sterilization is the validated method for sterilizing Omni-Tract products in a wrapped tray. No other methods have been validated and approved for use with Omni-Tract products. NOTE: Instruments, regardless of sterilization cycle used, will remain too hot for immediate use. Allow sufficient dry time and cooling before use. Prevacuum: Utilize one of the Prevacuum sterilization cycles listed in Table 2 to ensure the Omni-Tract Surgical Retractor Systems are properly sterilized after use. Table 2: Prevacuum Sterilization Cycle Parameters Cycle Type Temperature Exposure Time Dry Time (minimum) Prevacuum 270F (132C) 4 minutes 90 minutes Prevacuum 273F (134C) 3 minutes 110 minutes Prevacuum 273F (134C) 18 minutes 120 minutes 4. Complete the Acknowledgement Form included in Ap
Initiated
2026-06-05
Root cause
Labeling design
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brunei Darussalam, Bulgaria, Canada, Chile, China, Czechia, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea (the Republic of), Kuwait, Lithuania, Malaysia, Malta, Mauritius, Mexico, Nepal, Netherlands (the), New Zealand, Norway, Pakistan, Palestine, State of, Philippines (the), Poland, Portugal, Puerto Rico, Qatar, Republic of North Macedonia, R¿union, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan (Province of China), Thailand, Trinidad and Tobago, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland, (the), Viet Nam.
Recall number
Z-2714-2026
Recall status
Open, Classified
CDRH record ID
221034
FDA classified
2026-07-09
Recall event number
99193

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗