Z-2715-2020: Proteus 235 The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medi

Recall record Z-2715-2020 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2715-2020
Company in FDA record
Ion Beam Applications S.A.
FDA record creation date
Jul 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

IBA became aware of an issue to resume aborted irradiations because the plan to complete the treatment cannot be loaded.

Source facts

Action
An Urgent Medical Device Correction notification will be communicated to all impacted IBA customers (users of the device), starting July 23, 2020. This letter instructs the customer: If this situation were to occur, it is recommended to complete the aborted treatment in Standalone Mode (please refer to the steps described in annex I and to the part Using adaPTdeliver in Standalone Mode of the Proton Therapy System Clinical Users Guide). IBA will modify the software to remove conditions that make impossible to load the plan to complete the aborted treatment. The solution will be deployed on your site by June 30, 2022.
Initiated
2020-07-17
Root cause
Software design
Terminated
2021-08-30
Distribution
US: VA OUS: Belgium
Recall number
Z-2715-2020
Recall status
Terminated
CDRH record ID
182609
FDA record created
2020-07-27
Recall event number
86076

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗