IBA became aware of an issue to resume aborted irradiations because the plan to complete the treatment cannot be loaded.
Source facts
Action
An Urgent Medical Device Correction notification will be communicated to all impacted IBA customers (users of the device), starting July 23, 2020. This letter instructs the customer: If this situation were to occur, it is recommended to complete the aborted treatment in Standalone Mode (please refer to the steps described in annex I and to the part Using adaPTdeliver in Standalone Mode of the Proton Therapy System Clinical Users Guide). IBA will modify the software to remove conditions that make impossible to load the plan to complete the aborted treatment.
The solution will be deployed on your site by June 30, 2022.
Initiated
2020-07-17
Root cause
Software design
Terminated
2021-08-30
Distribution
US: VA
OUS: Belgium
Recall number
Z-2715-2020
Recall status
Terminated
CDRH record ID
182609
FDA record created
2020-07-27
Recall event number
86076