Z-2716-2024: Philips Allura Xper systems (R8.1 with Poly-G Stand) Interventional Fluoroscopic X-Ray System, Model Numb

Recall record Z-2716-2024 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2716-2024
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Aug 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A half-threaded bolt was included in the LTE kits instead of the correct full-threaded bolt.

Source facts

Action
An URGENT Medical Device Correction notification letter dated 7/31/24 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients. a. Affected systems may continue to be used in accordance with the Instructions for Use (IFU). b. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure the letter is in a place likely to be seen/viewed. c. Circulate this notice to all users of the system so that they are aware of the issue. d. Should you experience uncontrolled C-arm movement, call your local Philips representative to report the event. e. Return the attached response form to Philips promptly and no later than 30 days from receipt to confirm that the users of the system have reviewed and understood this Medical Device Correction and required actions to be taken. Actions planned by Philips IGT-S to correct the problem Philips will contact all affected customers to arrange for a Field Service Engineer visit to install an updated Lifetime Extension Kit with the correct full-threaded bolt (reference FCO72200590). If you need any further information or support concerning this issue, contact your local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center).
Initiated
2024-08-08
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of IL and OR. The countries of Hungary, Ireland, Poland, UK.
Recall number
Z-2716-2024
Recall status
Open, Classified
CDRH record ID
209596
FDA classified
2024-08-22
Recall event number
95144

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗