Z-2717-2011: iView DAB Detection Kit, Part number: 760-091 (05266157001) This detection kit is intended for in vitro

Recall record Z-2717-2011 for Ventana Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2717-2011
Company in FDA record
Ventana Medical Systems Inc
FDA classification date
Jul 1, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated because Ventana Medical Systems, Inc. (Ventana), has identified the presence of precipitate in some DAB chromogen dispensers in two lots of ultraView Universal DAB Detection Kit (part number 760-500, lot A05885 and lot B00988) and one lot of iView DAB Detection Kit (part number 760-091, lot B01057A.)

Source facts

Action
The firm, Ventana, sent two letters dated May 16, 2011, an "URGENT Medical Device Recall" notice to customers who received the affected lots and an "URGENT Medical Device Correction" to customers of unaffected lots. The letters described the product, problem and actions to be taken. The customers were instructed to discontinue use of any kits from the lots identified above; follow their internal standard operating procedures for assessing potential impact to reported patient results, If their laboratory had already used kits from these lots; conduct visual inspections of any DAB chromogen dispensers in use as part of their daily maintenance routine; inspect their current inventory and contact their local customer support team (800-227-2155) to replace any unused or partially used kits from the affected lots. Ventana highly recommends use of same-slide controls, especially when testing for targets of therapy, to mitigate any impact on patient results. If you find a dispenser containing precipitate in lots not identified in this letter or have any questions and concerns, please notify/contact your local customer support team (800-227-2155) .
Initiated
2011-05-16
Root cause
Pending
Terminated
2012-02-10
Distribution
Worldwide distribution: USA (Nationwide) and countries of Canada, Germany and Israel.
Recall number
Z-2717-2011
Recall status
Terminated
CDRH record ID
100423
FDA classified
2011-07-01
Recall event number
58855

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗