Open original FDA record ↗Recall record
- FDA identifier
- Z-2717-2014
- Company in FDA record
- KCI USA, Inc.
- FDA classification date
- Sep 29, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Blade can shift during shipping which makes the product ineffective.
Source facts
- Action
- KCI sent an Urgent - Voluntary Medical Device Field Notification letter dated June 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is sending Field Notifications to affected customers instructing them to complete appropriate actions. Option I: Contact your local KCI Representative or KCI Customer Technical Support at 1-800-275-4524 and select Option 3 to arrange for return and replacement of the affected harvester(s). Option 2: If you prefer, you may re-set the blade after removal from the packaging by gently tapping the curved end of the harvester against the palm of your hand, as shown in Figure 3, until the blade is no longer visible in the harvester top plate, as shown in Figure 4.
- Initiated
- 2014-05-19
- Root cause
- Device Design
- Terminated
- 2015-09-18
- Distribution
- US nationwide Distribution
- Recall number
- Z-2717-2014
- Recall status
- Terminated
- CDRH record ID
- 127360
- FDA classified
- 2014-09-29
- Recall event number
- 68329