Skip to content

Public FDA record

Z-2717-2026: Brand Name: Versius Surgical System US Product Name: Vesius Surgeon Console US Model/Catalog Number: V-SS

Recall record Z-2717-2026 for CMR SURGICAL LIMITED. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2717-2026
Company in FDA record
CMR SURGICAL LIMITED
FDA classification date
Jul 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Secure Boot was mistakenly not enabled at manufacturing time, which presents a potential cybersecurity risk.

Source facts

Action
On June 16, 2026, the firm began notifying affected customers. The firm or its local distributor will proactively carry out the secure boot configuration of the affected consoles. ***Update*** On July 8, 2026, the firm issued an updated letter which clarified the potential risk and instructions for the user. Users were instructed to keep the device in a secure environment with restricted access whether it was in clinical use or not. Users should inspect the system for signs of damage or interference before each use. Tamper-evident labels are located on the casework and will become damaged in the event of unauthorized access. Do not use the system if there is any damage or visible signs of interference on the tamper-evident labels and report immediately to CMR Surgical directly on regulatory.external@cmrsurgical.com or via your local representative.
Initiated
2026-06-09
Root cause
Software design (manufacturing process)
Distribution
US Nationwide distribution in the state of California.
Recall number
Z-2717-2026
Recall status
Open, Classified
CDRH record ID
221187
FDA classified
2026-07-09
Recall event number
99280

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗