- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
This recall has been initiated due to K-wires with a distal tip diameter that may be greater than specified causing an interference fit with the canulated screw. Removing the K-wire during a procedure may dislocate the implant resulting in the implant loosening or being pulled out or position. Use of this product may result in a delay of the surgical procedure and/or pull-out of the facet screw
Source facts
- Action
- Interventional Spine sent an RECALL NOTIFICATION letter dated March 23, 2011, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. The customer was instructed to take the following actions: Immediately examine inventory and quarantine product subject to recall. In addition, if this product was further distributed, please notify the Vice President Quality Assurance at once. Consignee were asked to also complete and return the enclosed response form as soon as possible. Consignee with questions were directed to the Consignees' Regional Director or the Vice President Quality Assurance. For further questions, please call (949) 472-0006
- Initiated
- 2011-03-23
- Root cause
- Pending
- Terminated
- 2013-02-14
- Distribution
- Nationwide Distribution including the states of to NC, NY, TN & VA
- Recall number
- Z-2718-2011
- Recall status
- Terminated
- CDRH record ID
- 99407
- FDA classified
- 2011-07-01
- Recall event number
- 58465