Z-2718-2016: ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of p

Recall record Z-2718-2016 for ARJOHUNTLEIGH POLSKA Sp. z.o.o.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2718-2016
Company in FDA record
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
FDA record creation date
Oct 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ArjoHuntleigh has concluded that a device tipping hazard may occur on the current Sara Combilizer lifting mechanism, providing a potential for risk to the patient or the caregiver. The tipping phenomenon has been found during internal device testing. To date, there have been no adverse events or customer complaints related to the tipping phenomenon. The tipping hazard may occur when fully loaded Sara Combilizer in "stretcher mode" is lifted to the highest position (approximately 984 mm = 38 3/4") and tilted backwards to the angle of 25 degrees-28 degrees. Although such scenario has been defined as highly unlikely to occur during use with patients, as a commitment to quality, ArjoHuntleigh is providing customers with information on action needed. The Sara Combilizer may stay in use until the upgrade will be performed, on the condition that the tilt angle in the device ''stretcher mode" will be no greater than minus 15 degrees and the device is not loaded over 330 lbs (150 kg).

Source facts

Action
A Field Safety Notice, a Customer Response Form and the Sara Combilizer Addendum to the Instructions For Use (IFU) will be mailed to all affected customers.
Initiated
2016-06-29
Root cause
Device Design
Terminated
2019-03-08
Distribution
CO, AL, MA, NC, MI, NY, FL, MD, CA, VA, CT, TX, and IL; Australia, Austria, Denmark, Germany, Israel, Japan, Norway, Sweden, Switzerland, UAE, and UK.
Recall number
Z-2718-2016
Recall status
Terminated
CDRH record ID
147789
FDA record created
2018-10-11
Recall event number
74506

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗