Skip to content

Public FDA record

Z-2719-2011: PERPOS PLS SYSTEM, CATALOG NO. 9045-02 A sterile, medical device kit consisting of the components liste

Recall record Z-2719-2011 for Interventional Spine Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2719-2011
Company in FDA record
Interventional Spine Inc
FDA classification date
Jul 1, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

This recall has been initiated due to K-wires with a distal tip diameter that may be greater than specified causing an interference fit with the canulated screw. Removing the K-wire during a procedure may dislocate the implant resulting in the implant loosening or being pulled out or position. Use of this product may result in a delay of the surgical procedure and/or pull-out of the facet screw

Source facts

Action
Interventional Spine sent an RECALL NOTIFICATION letter dated March 23, 2011, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. The customer was instructed to take the following actions: Immediately examine inventory and quarantine product subject to recall. In addition, if this product was further distributed, please notify the Vice President Quality Assurance at once. Consignee were asked to also complete and return the enclosed response form as soon as possible. Consignee with questions were directed to the Consignees' Regional Director or the Vice President Quality Assurance. For further questions, please call (949) 472-0006
Initiated
2011-03-23
Root cause
Pending
Terminated
2013-02-14
Distribution
Nationwide Distribution including the states of to NC, NY, TN & VA
Recall number
Z-2719-2011
Recall status
Terminated
CDRH record ID
99408
FDA classified
2011-07-01
Recall event number
58465

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗