Z-2720-2024: Ion Endoluminal System, REF: 380748-65

Recall record Z-2720-2024 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2720-2024
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Aug 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.

Source facts

Action
On July 12, 2024, Intuitive issued a "Urgent Medical Device Correction" notification to affected consignees via E-Mail. 1. This notice needs to be passed on to all those who need to be aware within your organization or functions where the potentially affected devices have been transferred. 2. Please retain a copy of this letter, place a copy with your affected system, and keep the acknowledgement form for your files. 3. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 4. An Intuitive Representative will schedule a site visit to perform replacement of the affected FIM assembly. Once replaced by the Intuitive Representative the Ion System can be used as designed. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, regular mail, or fax.
Initiated
2024-07-12
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of AZ and CA.
Recall number
Z-2720-2024
Recall status
Open, Classified
CDRH record ID
209213
FDA classified
2024-08-23
Recall event number
95046

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗