Z-2722-2026: da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15

Recall record Z-2722-2026 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2722-2026
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Jul 10, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result in procedure delay or conversion to an alternate minimally invasive approach. Retrospectively reported recall.

Source facts

Action
On 5/12/2025, correction notices were emailed to customers informing them of the following: 1) Conduct pre-use inspection steps outlined prior to each procedure, see below. 2) If you encounter a drape that does not pass the inspection steps, contact firm customer service. 3) Complete and return the acknowledgement form via email to Recalls@intusurg.com or EU.fsca@intusurg.com If you need further information or support concerning this Safety Notice, please contact your Clinical Sales Representative or contact firm Customer Service at the numbers listed below: - North and South America: (800) 876-1310, Option 3 or mail: customerservice@intusurg.com. - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 or eucs@intusurg.com - South Korea: 02-3271-3200or support.korea@intusurg.com - Japan: 0120-56-5635 or 03-5575-1362 or csjapan@intusurg.com " Taiwan: +0800-86-8181 Pre-Use Inspection Steps: 1. Fully drape the da Vinci SP system according to user manual. 2. Press and hold the Instrument Arm Clutch button. Top Instrument arm clutch, not recommended. 3. While holding the Instrument Arm Clutch button, manually rotate the instrument drive cluster in one direction, 360-degrees. Then, rotate the instrument drive at least 360-degrees in the other direction. Stop immediately and press "deploy for draping" again to rehome the instrument drive cluster to replace the drape if you observe any of the following behaviors: a. Inability to rotate the instrument drive 360-degrees in both directions b. VSC message that drape is not fully attached (VSC message: "Check drape is secure to all 4 instrument arm clips" or "Insecure or missing drape. The drape may not be properly secured and could fall. Ensure the drape is properly connected to all 4 mounts to continue"). c. Drape outer ring falls off silver section of system. d. Any of the 4 dark gray marks disengage from the dark gray connectors during the rotation.
Initiated
2025-05-12
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of NJ, OR, NY, FL, AL, SD, TX, VA, WV, PA, CA, WA, CT, MA, IL, OK, OH, MI, NC, MO, KY, WI, AZ, CO, GA, MD, SC, IN, NV, TN, MS, KS, IA, UT, MN, LA, NE, ME and the countries of Germany, Sweden, United Kingdom, Italy, Belgium, Switzerland, Denmark, Japan, South Korea, Hong Kong, Taiwan, France, Ireland, Spain.
Recall number
Z-2722-2026
Recall status
Open, Classified
CDRH record ID
220086
FDA classified
2026-07-10
Recall event number
98890

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗