Z-2723-2026: ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infu

Recall record Z-2723-2026 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2723-2026
Company in FDA record
ICU Medical, Inc.
FDA classification date
Jul 10, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.

Source facts

Action
ICU Medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 06/10/2026 via hand delivered letter. The notice explained the issue, potential risk, and requested the following: "Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification. If the pump is in the state shown in Figures 1 and 2, do not press the Start button. Wait for the pump to complete the Cassette Test before taking further action. It is important to note that if the clinician waits for the Cassette Test to complete (and the START button is only pressed once), this potential scenario does not occur. If the pump displays a green background on the line with active animation indicating delivery is occurring and the VTBI is not decrementing, stop delivery and restart the line to resolve this issue. Complete and return the attached Customer Response Form to ICUMedMarketActionsTeam@icumed.com within 10 days of receipt to acknowledge your understanding of this notification." For further inquiries: For any questions regarding this action: https://icumed.custhelp.com/app/market-action. Additional technical information or assistance: Technical Support, 1-(800)-241-4002, option 3 or tsc.support@icumed.com. To report adverse events or product complaints: Global Complaint Management: 1-844-654-7780 (M-F, 8:00 am 5:00 pm CT) or ProductComplaintsPP@icumed.com.
Initiated
2026-06-10
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall number
Z-2723-2026
Recall status
Open, Classified
CDRH record ID
221022
FDA classified
2026-07-10
Recall event number
99192

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗