Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.
Source facts
Action
Transonic sent an Urgent Medical Device Recall letters and Response Forms dated July 27, 2015 to customers via Certified Mail, email and/or fax. The letter identified the affected product, problem and actions to be taken. Customers are instructed to contact Transonic Systems' Customer Service at 1-800-353-3569 or 1-607-257-5300 to arrange for replacement (repair/exchange) immediately.
Initiated
2015-07-27
Root cause
Software Manufacturing/Software Deployment
Terminated
2016-07-18
Distribution
Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.
Recall number
Z-2727-2015
Recall status
Terminated
CDRH record ID
139364
FDA classified
2015-09-03
Recall event number
71907