The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
Source facts
Action
Hospira sent an Urgent Device Field Correction letter, dated February 6, 2013, to customers. Customers were requested to return a reply form to Stericycle indicating that they have received the letter, and notified their users and/or customers. The customers were asked to supply the number of Plum A+ infusers at their facility. Product will be corrected by Hospira in the field.
Initiated
2013-02-06
Root cause
Nonconforming Material/Component
Terminated
2016-11-14
Distribution
United States nationwide distribution including PR, and USVI. Costa Rica, Turks and Caicos, and Dominican Republic.
Recall number
Z-2729-2016
Recall status
Terminated
CDRH record ID
148269
FDA classified
2016-09-06
Recall event number
64261