Skip to content

Public FDA record

Z-2729-2026: Orthofix MINIRAIL, STANDARD LENGTHENER Model/Catalog Number: M101, Unilateral orthopedic/trauma external

Recall record Z-2729-2026 for Orthofix U.S. LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2729-2026
Company in FDA record
Orthofix U.S. LLC
FDA classification date
Jul 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111

Source facts

Action
On June 8, 2026, Orthofix issued a Urgent Medical Device Recall Notification, via E.Mail. Orthofix asked consignees to take the following actions: 1.Discontinue use of the affected product lots immediately. 2.Immediately examine your inventory and quarantine any affected product and return them to Orthofix. 3. Complete the attached response form and return it to Orthofix. 4. This notice should be shared with all relevant personnel within your organization and forwarded to any other organization or facility where potentially affected devices may have been transferred.
Initiated
2026-06-10
Root cause
Process change control
Distribution
US: AZ CA CO CT FL GA ID IL IN LA MD MI MO MS MT NC NY OH OR SC SD TX UT VA WA WI OUS: None
Recall number
Z-2729-2026
Recall status
Open, Classified
CDRH record ID
221091
FDA classified
2026-07-14
Recall event number
99225

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗