Open original FDA record ↗Recall record
- FDA identifier
- Z-2730-2017
- Company in FDA record
- Beckman Coulter Inc
- FDA classification date
- Jul 13, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.
Source facts
- Action
- Recall notices will be sent via mail or email to customers the week of 05/15/2017. Customers are instructed to discard the affected calibrator and request a replacement. Call Client Services at (800) 526-3821 (US only)
- Initiated
- 2017-05-15
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-09-09
- Distribution
- US Nationwide Distribution in the states of AL, AR, AK, CA, CO, DE, FL, GA, ID, IL, IN, IO, KS, KY, LA, MA, MD, MS, MI, ME, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WV, WI, and Puerto Rico Brunei, Canada, China, Czech Republic, France, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Lebanon, Mexico, Netherlands, Panama, Paraguay, Philippines, Qatar, South Africa, Taiwan, Thailand, Turkey, United Arab Emirates, & United Kingdom.
- Recall number
- Z-2730-2017
- Recall status
- Terminated
- CDRH record ID
- 155844
- FDA classified
- 2017-07-13
- Recall event number
- 77364