Skip to content

Public FDA record

Z-2731-2011: DYONICS POWER Shaver Blades, 4.5mm Long Curved Full Radius Convex, package of 3. Catalog Number: 7205336

Recall record Z-2731-2011 for Smith & Nephew, Inc. Endoscopy Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2731-2011
Company in FDA record
Smith & Nephew, Inc. Endoscopy Division
FDA classification date
Jul 1, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Product has insufficient weld between the shaft tube and tip, could result in the tip separating from the shaft tube during use.

Source facts

Action
Smith & Nephew notified customers by an "URGENT - PRODUCT RECALL 1st NOTIFICATION" letter dated May 6, 2011. The letter describes the product, problem, and actions to be taken. The letter asked customers to call 800-343-5717 (option 3) or send an email to endo.andreturn@smith-nephew.com for a Return Authorization (RA) Number to return product, obtain credit or order replacement blades. Customers are to complete the form and return to the address provided being sure to reference the RA number. For any questions regarding this recall customers were instructed to call 508-337-3731.
Initiated
2011-05-06
Root cause
Process control
Terminated
2012-05-09
Distribution
Worldwide Distribution -- USA (nationwide) and countries of: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Germany, Hong Kong, Italy, Japan, Mexico, Netherlands, Turkey, and United Kingdom.
Recall number
Z-2731-2011
Recall status
Terminated
CDRH record ID
100187
FDA classified
2011-07-01
Recall event number
58733

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗