Skip to content

Public FDA record

Z-2731-2015: Temporary Total Artificial Heart (TAH-t) with Freedom Driver System REF 595000-001 Product Usage: Free

Recall record Z-2731-2015 for SynCardia Systems Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2731-2015
Company in FDA record
SynCardia Systems Inc.
FDA classification date
Sep 17, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

SynCardia Systems, Inc. is recalling Freedom Drivers due to a specific component of the drive mechanism may fail and cause the drive mechanism to stop pumping; resulting in the loss of life-sustaining function.

Source facts

Action
SynCardia sent an Urgent Medical Device Recall letter dated August 6, 2015 to affected customers. The letter identified the purpose, reason for the voluntary recall, risk to health, how to recognize that the device may fail, actioms to be taken by the customer/user, product and distribution information and type of action by the company. Customer were instructed to completed and return Recall Acknowledgement and Receipt Form to SynCardia. A SynCardia Clinical Support Specialist or other SynCardia representative will assist customers with replacement product and returning affected product. Contact Information For Questions Mary Pat Sloan Senior Vice President Global Certification and Logistics SynCardia Systems, Inc 1992 E Silverlake Rd Tucson, AZ 85713 Direct 520-547-7470 Toll-free 866-480-1122 Ext 1308 Monday-Friday 7:30 AM to 5:00 PM Pacific/Arizona time Hotline 24/7 866-771-9437 www.syncardia.com
Initiated
2015-08-06
Root cause
Component design/selection
Terminated
2015-11-04
Distribution
Worldwide Distribution -- US Nationwide in the states of AR, NC, CA, OH, PA, TX, MI, and VA; and the countries of Germany, Turkey and France.
Recall number
Z-2731-2015
Recall status
Terminated
CDRH record ID
139311
FDA classified
2015-09-17
Recall event number
71905

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗