Product has insufficient weld between the shaft tube and tip, could result in the tip separating from the shaft tube during use.
Source facts
Action
Smith & Nephew notified customers by an "URGENT - PRODUCT RECALL 1st NOTIFICATION" letter dated May 6, 2011. The letter describes the product, problem, and actions to be taken. The letter asked customers to call 800-343-5717 (option 3) or send an email to endo.andreturn@smith-nephew.com for a Return Authorization (RA) Number to return product, obtain credit or order replacement blades. Customers are to complete the form and return to the address provided being sure to reference the RA number. For any questions regarding this recall customers were instructed to call 508-337-3731.
Initiated
2011-05-06
Root cause
Process control
Terminated
2012-05-09
Distribution
Worldwide Distribution -- USA (nationwide) and countries of: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Germany, Hong Kong, Italy, Japan, Mexico, Netherlands, Turkey, and United Kingdom.
Recall number
Z-2732-2011
Recall status
Terminated
CDRH record ID
100188
FDA classified
2011-07-01
Recall event number
58733