Z-2732-2017: Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655

Recall record Z-2732-2017 for Nova Biomedical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2732-2017
Company in FDA record
Nova Biomedical Corporation
FDA record creation date
Jul 13, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test results.

Source facts

Action
Nova Biomedical issued Urgent Field Correction Notice on June 19, 2017, to all affected customers. The letter advises users of the problem and Nova Biomedical or an authorized Nova Biomedical representative will schedule a no charge update of your Stat Profile Prime Analyzer software to version VI1.0.883.0. A representative will contact customers to schedule this update. Questions: contact Nova Biomedical Technical Support at 1-800-545-6682, in Canada contact us at 1-800-263-5999, or outside of N. America contact your authorized Nova Biomedical subsidiary or distributor. For questions regarding this recall call 781-647-3700.
Initiated
2017-06-19
Root cause
Software design
Terminated
2018-08-14
Distribution
Worldwide Distribution - US (Nationwide) Foreign: Albania Algeria Argentina Australia Austria Bangladesh Benin Bolivia Brazil Canada Chile China Colombia Costa Rica Czech Republic ECUADOR Egypt El Salvador Finland France Germany Ghana Greece Guatemala Guyana Honduras Hong Kong Hungary ICELAND India Iran Israel Italy Ivory Coast Japan Jordan Kenya Korea Latvia Macedonia Malaysia Mexico Nepal Netherlands Nigeria Pakistan Pakistan Palestine Peru Philippines Poland Puerto Rico Republic of Panama Republica Dominicana Russia Saudi Arabia Slovak Republic Slovenia South Africa Sri Lanka Switzerland Taiwan Thailand The Netherlands Trinidad,W.I. Tunisia Turkey UK Uruguay Vietnam
Recall number
Z-2732-2017
Recall status
Terminated
CDRH record ID
156306
FDA record created
2017-07-13
Recall event number
77532

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗