Open original FDA record ↗Recall record
- FDA identifier
- Z-2732-2026
- Company in FDA record
- Orthofix U.S. LLC
- FDA classification date
- Jul 14, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111
Source facts
- Action
- On June 8, 2026, Orthofix issued a Urgent Medical Device Recall Notification, via E.Mail. Orthofix asked consignees to take the following actions: 1.Discontinue use of the affected product lots immediately. 2.Immediately examine your inventory and quarantine any affected product and return them to Orthofix. 3. Complete the attached response form and return it to Orthofix. 4. This notice should be shared with all relevant personnel within your organization and forwarded to any other organization or facility where potentially affected devices may have been transferred.
- Initiated
- 2026-06-10
- Root cause
- Process change control
- Distribution
- US: AZ CA CO CT FL GA ID IL IN LA MD MI MO MS MT NC NY OH OR SC SD TX UT VA WA WI OUS: None
- Recall number
- Z-2732-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221094
- FDA classified
- 2026-07-14
- Recall event number
- 99225