Open original FDA record ↗Recall record
- FDA identifier
- Z-2733-2011
- Company in FDA record
- St Jude Medical CRMD
- FDA classification date
- Jul 5, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated because St. Jude Medical made new software available for the SJM Confirm DM2100 implantable cardiac monitor (ICM) that incorporates enhancements to the devices sensing algorithm. These enhancements are intended to provide improved sensitivity and specificity for diagnosing arrhythmias. The user is presented with the option to upgrade new software into the SJM Confirm ICM
Source facts
- Action
- St. Jude Cardiac Rhythm Management Division sent an "Important Product Correction" letter on March 25, 2011 to all affected customers. The letter explains the reason for the recall and provided recommendations including a software upgrade by SJM Field Clinical Engineer or Technical Services. Customers were recommended to fill out and return a "Physician Acknowledgement Form". For additional information please contact your local St.Jude Medical Sales Representative, Field Clinical Engineer or the Technical Services Department at 800-722-3774
- Initiated
- 2011-03-31
- Root cause
- Software change control
- Terminated
- 2011-07-19
- Distribution
- Worldwide distribution: (USA) Nationwide including the States of CA, WA, FL, WV, NY, MD and LA, and the country of Germany
- Recall number
- Z-2733-2011
- Recall status
- Terminated
- CDRH record ID
- 99926
- FDA classified
- 2011-07-05
- Recall event number
- 58686