Skip to content

Public FDA record

Z-2733-2014: Array LaserLink, Manufactured by Lumenis, The Array" LaserLink" is a laser system accessory intended f

Recall record Z-2733-2014 for Lumenis, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2733-2014
Company in FDA record
Lumenis, Inc.
FDA classification date
Sep 30, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lumenis initiated a field-correction for the Array Laser Link", GA-0006700 (SN XXYYZZ) because the system may project an erroneous pattern display on the retina, which is different than the desired pattern.

Source facts

Action
Lumenis sent a Safety Advisory Notice dated July 15, 2014, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed that Lumenis would contact consignees to schedule a product update, to be completed at a convenient time. For questions consignees should call 801-656-2549. For questions regarding this recall call 801-656-2690.
Initiated
2014-08-15
Root cause
Software design
Terminated
2014-10-21
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to JAPAN, CHINA, and CANADA.
Recall number
Z-2733-2014
Recall status
Terminated
CDRH record ID
129733
FDA classified
2014-09-30
Recall event number
69139

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗