Z-2734-2020: Medfusion Syringe Pump Model 4000; Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1

Recall record Z-2734-2020 for Smiths Medical ASD Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2734-2020
Company in FDA record
Smiths Medical ASD Inc.
FDA classification date
Aug 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Inaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.

Source facts

Action
Consignees are being contacted via direct mailing and email. Consignees are asked to review and complete the Recall Response form indicating understanding of notification. Distributors are asked to forward the notification to parties to whom product has been distributed. Smiths Medical will send out 2 additional notifications by email. Phone calls will be made to non-responders.
Initiated
2020-06-26
Root cause
Software design
Distribution
Worldwide distribution - United Arab Emirates, United States, Afghanistan, Albania, Australia, Belgium, Bermuda, Canada, China, Germany, Denmark, Great Britain, Hong Kong, Indonesia, Ireland, India, Korea, Kuwait, Lebanon, Mexico, Malaysia, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Saudi Arabia, Trinidad and Tobago
Recall number
Z-2734-2020
Recall status
Open, Classified
CDRH record ID
182138
FDA classified
2020-08-06
Recall event number
85912

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗