Z-2736-2024: Afinion 2 Analyzer, model #1116986, US refurbished, CLIA waived.

Recall record Z-2736-2024 for Abbott Diagnostics Technologies AS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2736-2024
Company in FDA record
Abbott Diagnostics Technologies AS
FDA classification date
Aug 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Five Afinion 2 (AF2) Analyzer were incorrectly market configured as Moderate Complexity and distributed to customers. Use of moderately complex assays by waived laboratories could result in erroneous results.

Source facts

Action
The recalling firm issued two different letters dated 6/17/2024 beginning 6/17/2024 via first class mail, one to the distributor and one to the end user customer. The Dear Distributor letter referred the distributor to the enclosed customer notification letter for details on the recall for their reference, as the distributor letter did not contain an explanation for why the recall was being conducted or the impact on patient results. The distributor letter identified the devices that were affected and provided the following actions to be taken: (1) Abbott Diagnostics Technologies AS will be sending a letter on their behalf to impacted customers within the scope of the serial numbers noted in the letter; and (2) Please acknowledge receipt of this letter within 5 business days after receipt using the enclosed distributor acknowledgement form. The distributor is to acknowledge they have received the notification letter and that if they encounter any questions regarding the recall, they will direct them to Abbott Technical Services. The customer letter provided an explanation as to the issue, the impact on patient results or operator safety, and the necessary actions to be taken, which included: (1) Complete and return the "Customer Required Action" form accompanying the letter; (2) If the devices listed have been forwarded to other laboratories, the customer is requested to inform them of the recall and provide a copy of this letter; (3) Retain the letter for their laboratory records. The Customer Required Action form reports that the customer acknowledged receipt of the notice, confirms that all areas where the product could be located have been checked, and lists a number of steps that have been verified with regard to whether they have the affected instruments, were they redistributed to another facility, send a replacement, etc.
Initiated
2024-06-17
Root cause
Reprocessing Controls
Distribution
US Nationwide distribution in the states of AZ, IN, MA, MN, and OH.
Recall number
Z-2736-2024
Recall status
Open, Classified
CDRH record ID
208893
FDA classified
2024-08-27
Recall event number
94910

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗