Kits contain Zimmer Biomet mixing bowls and spatulas, which were recalled due to seal nonconformities which present risk of sterility breach. Health consequences from use of product include clinically insignificant extension of surgery to find another readily available device.
Source facts
Action
On June 10, 2026, the firm began notifying customers via First-Class Mail and email. Customers were instructed to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
Initiated
2026-06-10
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of CA and UT.
Recall number
Z-2736-2026
Recall status
Open, Classified
CDRH record ID
221272
FDA classified
2026-07-15
Recall event number
99318