Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.
Source facts
Action
The firm sent out a written notification of the recall on 7/31/2015.
Initiated
2015-07-31
Root cause
Component design/selection
Terminated
2017-01-26
Distribution
NY, OH, CA, SC, NJ, AZ, VA, IN, FL, MN, OK, AR, MO, WI, GA, SD, IA, IL, TX
Recall number
Z-2738-2015
Recall status
Terminated
CDRH record ID
139413
FDA classified
2015-09-04
Recall event number
71936