Z-2738-2020: Centrella Smart+ Bed, Catalog No. P7900 - Product Usage: intended to be used to treat or prevent pulmonar

Recall record Z-2738-2020 for Hill-Rom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2738-2020
Company in FDA record
Hill-Rom, Inc.
FDA record creation date
Jul 30, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected beds may have screws installed that do not meet material tensile strength requirements. Failure of this screw could result in serious patient or user injury.

Source facts

Action
On June 30, 2020, the firm mailed Urgent Medical Device Correction Notice letters to affected customers. Customers received letters with Centrella Smart+ Beds or Progressa Beds in the subject line depending on which affected products they had received. Customers were informed of the potential safety issue regarding nonconforming screws used in the beds, which may cause the frame to fail and lead to serious injury to the patient. Customers were instructed to do the following: 1. Until your affected beds are updated, Hill-Rom is urging customers to follow the instructions outlined below: -Caution your users that when utilizing the bed articulation functions, if there is any binding during movement or deformation in the frame, to take the bed immediately out of service until the bed is updated. -Follow the safety information as outlined in the Instructions for Use each time when setting the brakes. If the brakes do not hold after following the instructions, immediately take the bed out of service until the bed is updated. -Caution your users that after placing the bed in steer mode and beginning transport, if there is any difficulty with aligning one of the casters in the trailing position and maintaining control, take the bed immediately out of service until the bed is updated. -Follow the Instructions for Use each time when raising a siderail to the latched position. If the siderails are not latching properly, take the bed immediately out of service until the bed is updated. 2. Complete the response form attached to the notification and return it to Hill-Rom. 3. Contact Hill-Rom at 800-445-3720 or us.service@hillrom.com if you have any questions. 4. Hill-Rom will be reaching out to you to assist with the planning of your bed(s) correction. Hill Rom will follow-up with a phone call if they have not received a reply within 10 days. Hill-Rom will contact you to schedule your correction. The correction will need to be completed without a patient
Initiated
2020-06-30
Root cause
Nonconforming Material/Component
Terminated
2022-10-20
Distribution
Worldwide distribution - US nationwide distribution including in the states of AR, AZ, CA, DE, IN, KY, MD, ME, MI, MO, NC, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV and the country of Canada.
Recall number
Z-2738-2020
Recall status
Terminated
CDRH record ID
182167
FDA record created
2020-07-30
Recall event number
85925

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗