Skip to content

Public FDA record

Z-2738-2024: DR 800. Digital Radiography X-ray System.

Recall record Z-2738-2024 for Agfa N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2738-2024
Company in FDA record
Agfa N.V.
FDA classification date
Aug 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the front lever chain of the DR 800 table to fail.

Source facts

Action
Consignees were sent an URGENT FIELD SAFETY NOTICE by email, dated 7/29/24. In the notice, consignees are asked to check the correct tightening of the floor plate fixing bolts if a high and persistent noise is heard from the device during the tilting or elevation movement of the tabletop. If the noise persists, the device must be stopped immediately in a horizontal position at the maximum height. Agfa will coordinate with consignees to schedule an inspection of affected devices and to install hardware intended to reinforce equipment before 10/15/24. The recall notification is to be shared within consignee organizations and forwarded to those devices were transferred to. Consignees are asked to maintain awareness of this notice until devices are corrected. Completed response forms are to be provided via email to cassandra.mcgowan@agfa.com. Consignees with any questions are to call 1-877-777-2432 and reference PRB2001106.
Initiated
2024-07-29
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, IL, IN, KS, LA, MA, MO, NJ, NC, OH, OK, SC, VA, WV, WI; and the countries of AU, BE, BW, CA, CO, CZ, DE, ES, FR, GB, GR, IL, IN, IT, KZ, MX, PL, RS, UZ, ZA.
Recall number
Z-2738-2024
Recall status
Open, Classified
CDRH record ID
209480
FDA classified
2024-08-27
Recall event number
95115

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗