Skip to content

Public FDA record

Z-2740-2011: NucliSens EasyQ Enterovirus serotype 68 v1.1: Catalog number: 200300 In Vitro Diagnostic Test

Recall record Z-2740-2011 for bioMerieux, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2740-2011
Company in FDA record
bioMerieux, Inc.
FDA classification date
Jul 6, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The NucliSENS EasyQ Enterovirus Reagents are unable to detect Enterovirus serotype 68, in respiratory samples.

Source facts

Action
BioMerieux Vitek, Inc. sent an "Urgent Product Correction Notice" letter on/about May 10, 2011. The letter included; a description of the product, problem, possible impact to the patient and recommendations. Recommendations included instructing customers to discontinue the use of the kit for the detection of Enterovirus 68 and Rhinovirus 87 serotypes and noted the corrected IFU will be available the first week of April 2011. An Acknowledgement form was attached to be returned via fax. For additional questions please contact your local BioMerieux Clinical Customer representative at (800) 682-2666, option 3, option 2.
Initiated
2011-05-10
Root cause
Device Design
Terminated
2011-10-06
Distribution
(USA) Nationwide distribution including the states of MD, MI, MN, MT, NY and PA
Recall number
Z-2740-2011
Recall status
Terminated
CDRH record ID
100303
FDA classified
2011-07-06
Recall event number
58792

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗