Skip to content

Public FDA record

Z-2741-2017: MAKO RIO Standard System, internal cPCI Card CageP/N: 201337. Intended to assist the surgeon in providin

Recall record Z-2741-2017 for Mako Surgical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2741-2017
Company in FDA record
Mako Surgical Corporation
FDA record creation date
Jul 18, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An intermittent electrical problem that could lead to a loss of system power due to a loose screw connection.

Source facts

Action
Customers were sent notices titled "Urgent Medical Device Correction Notification" on 06/12/2017 via UPS Next Day Mail. The firm notifies consignees of the reason for recall and instructs that users of the recalled product be provided the recall notice. Users may continue to use their MAKO system; and, Stryker/MAKO Service personnel will contact users o schedule time to inspect and replace the card cage, if necessary. Also, users are instructed to return the completed Business Reply Form within 5 business days. The completed form can be sent to strykerortho7391@stericycle.com or faxed to 855-544-4807. Any questions can be directed to Clayton Odor at 954-628-0502.
Initiated
2017-06-07
Root cause
Employee error
Terminated
2020-06-16
Distribution
Distributed to the states of AZ, CA, CO, CT, FL, GA, ID, LA, MA, MO, MS, NC, OH, PA, TN, TX, & WA.
Recall number
Z-2741-2017
Recall status
Terminated
CDRH record ID
156493
FDA record created
2017-07-18
Recall event number
77553

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.