Skip to content

Public FDA record

Z-2741-2018: CPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is

Recall record Z-2741-2018 for Compass Health Brands (Corporate Office). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2741-2018
Company in FDA record
Compass Health Brands (Corporate Office)
FDA classification date
Aug 20, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A May 2015 design change in the cushion seal replacement part and accompanying elbow replacement part resulted in incompatibility of the new cushion seal with the previous design of the elbow. Use of the new cushion seal with the previous design elbow could lead to unacceptable CO2 porting. As there was no model number change for this design change, customers may not be aware of the incompatibility.

Source facts

Action
On June 19. 2018, the firm sent Urgent Recall Field Correction notices via certified mail. The notice alerted customers to the following product issue: Compass Health Brands Corp. has recently identified the need to clarify compatibility information for the replacement cushion seals for the Model# PB7801 Zzz-Mask SG Series Full Face CPAP Mask. >- A design change made to the cushion seal replacement part (and accompanying elbow replacement part) in May 2015 (18 months prior to Compass Health acquisition of the Pro basics product line) resulted in incompatibility of the new cushion seal with the previous design of the elbow. >- While no complaints or injuries have been reported to-date, use of the new cushion seal with the previous design of elbow could result in unacceptable CO2 porting. >- Since a model number change did not accompany this design change, customers and their consumers would likely not be aware of this incompatibility. Customers were asked to do the following: Ensure all affected personnel are fully informed of this notice. " Forward this notice to your Regulatory Compliance Manager, Purchasing Manager, Customer Service Manager and Field Technicians. " Advise your customers regarding the proper combination of cushion and elbow. The attached consumer communication letter can be used to communicate this information. As noted on the attached consumer communication letter, any consumers who are currently using an incompatible cushion/elbow combination should be instructed to discontinue use of the cushion and elbow and should be instructed to contact their dealer to obtain replacement kits. " Dispose of any current stock you may have of the affected cushion model numbers (PB781S, PB781M, PB781L, 781S, 781M and 781L manufactured after May 2015) and contact Compass Health for replacement kits (NOTE: Full mask model numbers are NOT affected by this recall field correction). " Please contact Compass Health Brands Corp. Customer S
Initiated
2018-06-18
Root cause
Component change control
Terminated
2020-05-05
Distribution
US nationwide distribution to AL, CT, FL, GA, KY, MS, NC, NY, TN, and TX.
Recall number
Z-2741-2018
Recall status
Terminated
CDRH record ID
165713
FDA classified
2018-08-20
Recall event number
80545

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗