Skip to content

Public FDA record

Z-2741-2020: Alaris System PC Unit Model 8015. modular infusion pump and monitoring system

Recall record Z-2741-2020 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2741-2020
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Aug 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.

Source facts

Action
On 06/30/2020, Urgent Medical Device Recall Notifications were mailed to Directors of Biomedical Engineering, Nursing, Risk Management, Environmental Services. Actions for Clinical Users: This situation only occurs when the device is running on battery power. 1. If this issue occurs, immediately plug the PC unit into AC power and restart the infusions. 2. Obtain a replacement PC unit as soon as possible. 3. Ensure back-up devices are readily available when infusing critical medications. Please also ensure that you have back-up monitoring devices (e.g., EtCO2, SpO2) that are critically important to the patient's care. 4. Complete and return the Customer Response Card. Actions for Cleaning Personnel: Inspect all PC unit batteries for missing screws and washers during cleaning. Should missing screws or washers be identified, send PC unit to Biomedical Engineering for repair. Expedite the inspection of all PC unit batteries for loose or missing screws and washers. If loose or missing screws and washers are identified, please replace them and torque any loose screws or washers to 6 in/lb and perform battery conditioning. See Service Bulletin 592A. Contact recalling firm at 1-800- 482-4822 to order a package of screws and washers. Battery should be replaced every two years by qualified service personnel using authorized batteries. Actions by Recalling Firm: Battery Pack Assembly procedure was revised to add clarity for reassembly of the battery, including proper use of screws and washers. Updated Service Bulletin to be provided in August 2020. An initial quantity of replacement screws and washers will be provided for use. Future orders for replacement batteries will include screws and washers System rental providers, please provide a copy of this letter and enclosures to your customers who are renting affected devices. Recall resources/customer support: www.bd.com/alaris-system-hardware-recall Questions - Recall Support Center: 858-617-1316, GMB-Ala
Initiated
2020-06-30
Root cause
Nonconforming Material/Component
Terminated
2024-04-22
Distribution
U.S.: CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, KS, MN, MI, WA, UT, MD, TN, CO, NJ, NE, ND, WV, KY, IA, MS, AL, DE, AK, VA, OK, NM, GA, CT, SD, NV, ID, MT, ME, VT, WY, NH, MP, DC, GU, PR, AS, RI. O.U.S.: TR, QA, CA, AE, KW, MX, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, BH
Recall number
Z-2741-2020
Recall status
Terminated
CDRH record ID
182144
FDA classified
2020-08-06
Recall event number
85915

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗