Z-2741-2026: The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a steri

Recall record Z-2741-2026 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2741-2026
Company in FDA record
Maquet Cardiovascular, LLC
FDA classification date
Jul 15, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.

Source facts

Action
Getinge issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/18/2026 via FedEx. The notice explained the issue, risk to health, and requested the following actions be taken: " Immediately identify and quarantine all affected Vasoview Hemopro 2 devices from lot 3000547797. " Do not use affected devices. " If an affected device was used, no further intervention is required. Standard post-operative wound care per facility protocol should be followed. " If any affected devices were transferred within your organization or to another organization, ensure this notice is shared with the appropriate individuals. " Contact Getinge Customer Service 1-888-880-2874 or your Getinge representative to coordinate return of the affected units and replacement product. " Complete and return the enclosed response form to Getinge by e-mailing a copy to recallresponses.qrc@getinge.com.
Initiated
2026-06-02
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.
Recall number
Z-2741-2026
Recall status
Open, Classified
CDRH record ID
221161
FDA classified
2026-07-15
Recall event number
99259

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗