Z-2742-2016: ADVIA Centaur PSA Assay; Cat No. 06574155/SMN 10310292 (100 test)/02676506/SMN 10310293 (500 test)

Recall record Z-2742-2016 for Siemens Healthcare Diagnostics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2742-2016
Company in FDA record
Siemens Healthcare Diagnostics, Inc
FDA record creation date
Sep 7, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Communication provided to emphasize that the PSA values should be interpreted in accordance with current clinical guidelines for defining biochemical recurrence following radical prostatectomy.

Source facts

Action
An Urgent Medical Device Correction (UMDC) was sent to all affected Siemens Healthcare Diagnostics customers in the United States on July 29, 2016 to be delivered to customers on August 1, 2016. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on July 28, 2016. These emphasize that the prostate specific antigen (PSA) values should be interpreted in accordance with current clinical guidelines for defining biochemical recurrence following radical prostatectomy (e.g., the 2013 American Urological Association (AUA) Guidelines or the 2015 European Association of Urology (EAU)). These guidelines define biochemical recurrence of prostate cancer as a detectable or rising PSA value post-radical prostatectomy that is > or = 0.2 ng/mL (ug/L) with a second confirmatory level of > or = 0.2 ng/mL (ug/L). The communications also provide functional sensitivity from a recent study to give an expectation of precision at the low end of the assay.
Initiated
2016-08-09
Root cause
Under Investigation by firm
Terminated
2020-01-06
Distribution
Nationwide Albania Algeria Angola Armenia Austria Bahrain Belgium Bulgaria Burkina Faso Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France Georgia Germany Great Britain Greece Hungary Iceland Iraq Italy Jordan Kazakhstan Latvia Lithuania Luxembourg Macedonia Malta Netherlands Norway Pakistan Poland Portugal Qatar Romania Russian Fed. Senegal Serbia Slovakia Slovenia Spain Sweden Switzerland Tunisia Turkey Unit.Arab Emir. Uzbekistan Vatican
Recall number
Z-2742-2016
Recall status
Terminated
CDRH record ID
148778
FDA record created
2016-09-07
Recall event number
74917

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗