Open original FDA record ↗Recall record
- FDA identifier
- Z-2742-2016
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc
- FDA record creation date
- Sep 7, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Communication provided to emphasize that the PSA values should be interpreted in accordance with current clinical guidelines for defining biochemical recurrence following radical prostatectomy.
Source facts
- Action
- An Urgent Medical Device Correction (UMDC) was sent to all affected Siemens Healthcare Diagnostics customers in the United States on July 29, 2016 to be delivered to customers on August 1, 2016. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on July 28, 2016. These emphasize that the prostate specific antigen (PSA) values should be interpreted in accordance with current clinical guidelines for defining biochemical recurrence following radical prostatectomy (e.g., the 2013 American Urological Association (AUA) Guidelines or the 2015 European Association of Urology (EAU)). These guidelines define biochemical recurrence of prostate cancer as a detectable or rising PSA value post-radical prostatectomy that is > or = 0.2 ng/mL (ug/L) with a second confirmatory level of > or = 0.2 ng/mL (ug/L). The communications also provide functional sensitivity from a recent study to give an expectation of precision at the low end of the assay.
- Initiated
- 2016-08-09
- Root cause
- Under Investigation by firm
- Terminated
- 2020-01-06
- Distribution
- Nationwide Albania Algeria Angola Armenia Austria Bahrain Belgium Bulgaria Burkina Faso Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France Georgia Germany Great Britain Greece Hungary Iceland Iraq Italy Jordan Kazakhstan Latvia Lithuania Luxembourg Macedonia Malta Netherlands Norway Pakistan Poland Portugal Qatar Romania Russian Fed. Senegal Serbia Slovakia Slovenia Spain Sweden Switzerland Tunisia Turkey Unit.Arab Emir. Uzbekistan Vatican
- Recall number
- Z-2742-2016
- Recall status
- Terminated
- CDRH record ID
- 148778
- FDA record created
- 2016-09-07
- Recall event number
- 74917