Z-2743-2020: Remisol Advance running Software Version 1.12 The Normand Info Remisol Advance Data Manager collects and

Recall record Z-2743-2020 for Normand-Info S.A.S.U.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2743-2020
Company in FDA record
Normand-Info S.A.S.U.
FDA record creation date
Jul 31, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Results from repeated run for WBC (White Bloodcell Count), UWBC (Urine Whit Bloodcell Count) and PLT (Platelets) are deleted in RADV. Use may lead to erroneous results.

Source facts

Action
During the week of 04/27/2020, the Beckman Coulter notified affected customers via mail or email of " URGENT MEDICAL DEVICE RECALL". The letter indicated, if you are affected because your system includes one or more affected analyzers please ensure to order reruns through REMISOL and not through the analyzer until you receive the update. The resolution will be addressed by updating to the next version, version 1.12. You will be contacted by the company 's Field Associate to arrange a date and time for the upgrade to the new version. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you are a centralized license holder, please provide the other affected laboratories of your organization or association with a copy of this letter. So that the firm are assured you have received this important communication, please respond within 10 days in one of the following ways: - Electronically, if you received this communication via email. - Manually, complete and return the enclosed Response Form. Return completed form via email or fax at (714) 961-4234. If you have any questions regarding this notice, please contact the company via the firm's website or by phone by contacting your local representative. Firm telephone is (800) 526-3821.
Initiated
2020-04-30
Root cause
Software design
Terminated
2022-02-10
Distribution
Worldwide distribution: US (nationwide) including states of: CA, GA, IL, IN, KS, MI, MO, NY, NC, OH, PA, SD, TN, VA, and WV; and countries of: Bahrain, Belgium, Canada, China, Czech Republic, France, Germany, Hong Kong, Israel, Malaysia, Netherlands, Switzerland, Taiwan,Turkey, and United Kingdom.
Recall number
Z-2743-2020
Recall status
Terminated
CDRH record ID
181556
FDA record created
2020-07-31
Recall event number
85693

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗