Z-2743-2026: Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Recall record Z-2743-2026 for Daavlin Distributing Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2743-2026
Company in FDA record
Daavlin Distributing Company
FDA classification date
Jul 15, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

Source facts

Action
Daavlin Distributing Company initially notified its consignees about the issue beginning on 3/5/2026 via telephone following with by an email. The firm a follow up URGENT MEDICAL DEVICE Advisory notice on 4/9/2026 with the same information via US mail and email. On 7/10/2026 the firm sent an URGENT MEDICAL DEVICE RECALL notice to all non-responding consignees. The notice explained the issue, potential risk to health, and requested the following: "" Verify your device is operating ClearLink Controller Software Version 3.17 as follows: a) Ensure your machine is plugged in and on b) Locate and press the gear icon in the top right corner of the controller screen c) Select Machine Info d) Select Software Version e) Review Software Information " Confirm the serial number of your device is 123567890abc. The serial number can be found in all the following places: " For devices that remain in the original packaging, it can be found on the outside of the box " For devices that have been removed from the original packaging, it can be found on the back of your machine centered on the top of the Spine Cover or on the bottom of the Spine Cover "Once you confirm your device is identified as an affected device operating ClearLink Controller Software Version 3.17, contact Daavlin Technical Services immediately (see contact information below). Daavlin will place the device into Fixed Calibration Mode as an interim mitigation until ClearLink Controller Software Version 3.18 can be installed. " Complete the enclosed Recall Response Form and return it within 10 business days. Share this notice with all users of this device. " If you distributed the affected device to another facility, immediately provide them with a copy of this notice. " For questions, please contact: Daavlin Technical Services, Phone: (419) 633-3415, Email: rgolding@phothera.com
Initiated
2026-03-05
Root cause
Software Design Change
Distribution
US Nationwide distribution.
Recall number
Z-2743-2026
Recall status
Open, Classified
CDRH record ID
221313
FDA classified
2026-07-15
Recall event number
99331

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗