- Recall classification
- Class II
Reason stated in the source
Bubble Leak Testing on AbViser AutoValve IAP Monitoring Devices without Transducer confirmed the potential for a pinhole breach in the sterile barrier.
Source facts
- Action
- The firm, ConVaTec, sent an "URGENT: VOLUNTARY MEDICAL DEVICE RECALL" letter dated July 2, 2018, to its customers via United Parcel Service of America (UPS) Ground with tracking on July 3, 2018, with emails sent to distributors on July 9, 2018. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1. Inspect your inventory to confirm if you have any of the affected product codes. 2 Perform a count of affected product currently in inventory. Complete the enclosed response form and return it to the address on the response form. Return the attached Recall Response Form even if no affected product is in inventory. 3 If you have any affected product in inventory please contact ic@convatec.com or call the Customer Interaction Center at the toll free number 1-800-422-8811. You will be provided further instructions regarding how to return affected product and arrange credit.
- Initiated
- 2018-07-03
- Root cause
- Nonconforming Material/Component
- Terminated
- 2023-01-24
- Distribution
- worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Canada, Bahrain, Bangladesh, Belgium, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hungary, India, Israel, Italy, Kenya, Kuwait, Malta, Mauritius, Netherlands,Norway, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates, and Vietnam.
- Recall number
- Z-2744-2018
- Recall status
- Terminated
- CDRH record ID
- 165706
- FDA record created
- 2018-08-10
- Recall event number
- 80539