Open original FDA record ↗Recall record
- FDA identifier
- Z-2744-2020
- Company in FDA record
- Sutter Medizintechnik GmbH
- FDA record creation date
- Jul 31, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The label on the cardbox and the blister show the correct symbols for sterilization using irradiation and storage at 5 C to 40 C, the package insert (IFU) is labelled incorrectly. It shows the symbol for sterilization using Ethylene Oxide with the description for sterilization using irradiation.
Source facts
- Action
- Sutter Medizintechnik issues a recall letter via email dated July 2020. This letter instructs customers/distributors to do the following: Please further notify anyone you have distributed the product to. Please complete and return the acknowledgement and receipt form. Please exchange the IFU inside the package. You may follow the instruction as shown in the pictures. As a precautionary measure, do NOT use sharp tools such as forceps and do not perform the exchange with sharp fingernails.
- Initiated
- 2020-06-23
- Root cause
- Labeling design
- Terminated
- 2020-08-21
- Distribution
- US Nationwide distribution including in the states of MD, NJ, MI, NC, IN, CA, AL, WA, OK, MS, GA, FL, SC, TX, CO, MN, MO, KS.
- Recall number
- Z-2744-2020
- Recall status
- Terminated
- CDRH record ID
- 182545
- FDA record created
- 2020-07-31
- Recall event number
- 86035