Open original FDA record ↗Recall record
- FDA identifier
- Z-2745-2011
- Company in FDA record
- Barrx Medical, Inc.
- FDA classification date
- Jul 6, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Complaints of malfunction of the HALO ULTRA Ablation Catheter related to the device pivot mechanism's inability to return the electrode cap to a flat neutral position.
Source facts
- Action
- BARRX Medical, Inc. sent a "PRODUCT RECALL NOTIFICATION" letter dated May 25, 2011 via Federal Express to all of their customers. The letter described the product, problem and actions to be taken. The customers were instructed to discontinue use of the recalled product. BRRX Medical Territory Managers will be in contact with each facility and arrange for pick up of the recalled product. The "Steps for Voluntary Recall" form that is enclosed with each letter provides customers with complete instructions on segregating the product, completing the Tracking/Verification Form, obtaining Returned Good Authorization (RGA) number and preparing product for pick up. Customers are to complete and return the Tracking/Verification Form by fax to 408-738-1741 which should include amount of product on hand, contact name and hospital/distributor information. Questions should be directed to BRRX Medical Customer Service at 888-662-2779.
- Initiated
- 2011-05-25
- Root cause
- Component design/selection
- Terminated
- 2011-09-07
- Distribution
- Worldwide Distribution -- USA (nationwide) and the countries of the United Kingdom and China
- Recall number
- Z-2745-2011
- Recall status
- Terminated
- CDRH record ID
- 100503
- FDA classified
- 2011-07-06
- Recall event number
- 58908