Z-2745-2026: Power Over Fiber Base Module

Recall record Z-2745-2026 for Terex Utilities. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2745-2026
Company in FDA record
Terex Utilities
FDA classification date
Jul 30, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning statement

Source facts

Action
A letter was received stating the product, problem and actions to be taken. The following actions being taken include: A. Develop and carry out a process to remove incorrect labeling and install all labels required under IEC 60825-1 2014. B. Generate, detail & submit full CAP (Corrective Action Plan) for FDA approval C. Generate initial laser product report or supplemental as appropriate. D. Equipment update to be performed by Terex personnel at customer location E. Procedures: * Fully certify device as Class 1 under the guidelines of FDA Laser Notice #56 by applying the IEC 60825-1 2014 laser certification standard * Update and distribute compliant UI (User Information) * Update Sales information as appropriate * Generate and apply administrative and safety labeling * Formalized quality control procedures and record keeping. If you have any questions, contact Director of Engineering at +16058824000 or email: dan.brenden@terex.com
Initiated
2026-05-11
Root cause
Radiation Control for Health and Safety Act
Distribution
U.S. Nationwide Distribution and International Distribution
Recall number
Z-2745-2026
Recall status
Open, Classified
CDRH record ID
221451
FDA classified
2026-07-30
Recall event number
99382

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗