Open original FDA record ↗Recall record
- FDA identifier
- Z-2746-2016
- Company in FDA record
- Maquet Cardiovascular Us Sales, Llc
- FDA record creation date
- Sep 7, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Post Market Surveillance cases reported to Maquet showing instances in which the welded joint between the square mount and the first fixable swivel joint of Maquet's Connecting Bracket (Fixture) Device is broken.
Source facts
- Action
- Maquet sent an "Urgent Product Recall Medical Device Correction Letter and Medical Device Correction Response Form" dated August 12, 2016 via UPS to their affected customers. Customers were advised to inspect all devices that include the fixture for damage before every use. Customers were asked to fill out and sign the enclosed Medical Device Correction Response Form and return it via email to FieldAction@maquet.com or fax to 1-973-909-9927. Customers with questions were advised to contact their MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383. For questions regarding this recall call 973-709-7487.
- Initiated
- 2016-08-12
- Root cause
- Use error
- Terminated
- 2017-05-04
- Distribution
- Nationwide Distribution to CA, FL, IA, IN, MO, UT and WI
- Recall number
- Z-2746-2016
- Recall status
- Terminated
- CDRH record ID
- 148768
- FDA record created
- 2016-09-07
- Recall event number
- 74915