Z-2746-2018: MEVION S250/MEVION S250i Proton Beam Radiation therapy system Product Usage: Indicated for the delivery

Recall record Z-2746-2018 for Mevion Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2746-2018
Company in FDA record
Mevion Medical Systems, Inc.
FDA record creation date
Aug 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

QFix has advised Mevion Medical Systems that they are identifying unauthorized combinations of QFix Couch Tops with non-QFix devices in clinical use. The kVue Oen Touch latch consists of a precision latch mechanism that can be damaged when used with incompatible (third-party) accessories. This can create unsafe conditions for use and may compromise the subsequent use of any kVue inserts, whether authorized or not. Use of unauthorized or unvalidated inserts may lead to an incorrect patient treatment or physical injury.

Source facts

Action
On June 29, 2018, the firm sent out a User Notice to consignees to alert customers to the Advisory Notice issued by QFix regarding the use of an Orfit HP Pro Solution combined with the QFix kVue couch top. Customers using third-party inserts are advised to immediately inspect the latching mechanism for damage and contact the third-party insert vendor for advice on the use of the insert and confirmation of compatibility. Mevion will update user manuals to include the following warning about using alternative third party vendor inserts: WARNING: Interchangeable Couch Top inserts are integral to the immobilization, treatment beam planning, and physical safety of each patient. Use of unauthorized or unvalidated inserts may lead to an incorrect patient treatment or physical injury. Only use Mevion-provided accessories or devices that are confirmed by the third party suppliers to be compatible with the Mevion Couch for safe and effective patient treatment. Please contact Rob Cessac at rcessac@mevion.com or call 636-485-4962 for any questions or concerns.
Initiated
2018-06-29
Root cause
Device Design
Terminated
2018-10-02
Distribution
US Nationwide Distribution in the states of DC, FL, MO, NJ, OH, and OK.
Recall number
Z-2746-2018
Recall status
Terminated
CDRH record ID
165886
FDA record created
2018-08-11
Recall event number
80615

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗