Skip to content

Public FDA record

Z-2746-2020: SUTUREFIX ULTRA Suture, Product number 72203855

Recall record Z-2746-2020 for Smith & Nephew, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2746-2020
Company in FDA record
Smith & Nephew, Inc.
FDA record creation date
Jul 31, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A packaging error resulted in a XL drill being presented for use instead of the S drill. The XL drill is one inch longer than the S drill.

Source facts

Action
On June 24, 2020, the firm distributed Urgent Medical Device Recall Notices to their customers. Customers were informed that due to a packaging error, packages labeled as a 1.7mm twist drill (S) instead contained a 1.7mm (XL) drill. The overall length of the XL drill (12 inches) is longer than the S drill (11 inches). Customers were asked to do the following: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with the Urgent Medical Device Recall Notice, and quarantine them immediately. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. 2. If you have no product to return, please indicate this on the response form included with the letter. 3. If you have product to return, please list the item, batches and quantities that you are returning on the customer response form. 4. Complete the remainder of the customer response form, sign and send to FieldActions@smith-nephew.com. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
Initiated
2020-06-24
Root cause
Process control
Distribution
Domestic distribution to AL AR AZ CO GA IL IN KS KY MD MN MO NC NE NJ NY OK PA PR TN TX VA WI. Foreign distribution to Australia Austria Belgium Finland France Germany Great Britain Ireland Italy Japan Luxembourg Mexico Netherlands Poland Portugal Singapore South Africa Spain Sweden Switzerland
Recall number
Z-2746-2020
Recall status
Open, Classified
CDRH record ID
182372
FDA record created
2020-07-31
Recall event number
85981

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗