Z-2747-2018: Artis Q. angiography system. Model 10848280, 10848281, 10848282, 10848283, 10848353, 10848354 & 10848355

Recall record Z-2747-2018 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2747-2018
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Aug 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential hardware issue which may cause thermal effects and possible damage to system parts.

Source facts

Action
On June 22, 2018, Siemens Healthineers issued URGENT MEDICAL DEVICE CORRECTION notices to customers via courier service. In the event the problem occurs, the system can be operated normally, however, in the case of a leakage current, X-ray will not be possible. This may result in a situation where it is necessary to cancel or restart clinical treatment or to continue treatment on an alternative system. Please follow the standard emergency procedures you have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens will correct the error by updating all affected generators with a RCD (Residual Current Device) to detect leakage currents via Update Instruction AX060/16/S. Additionally, Siemens recommends installing a RCD (Residual Current Device) in the electrical system of the hospital as an appropriate backup system and general electrical protection. You may continue to use the Artis Q and Artis Pheno systems while awaiting implementation of the corrective action. Siemens Healthineers will contact customers to arrange a date to perform this corrective action. Please feel free to contact the service organization for an earlier appointment at 1-800-888-7436.
Initiated
2018-06-22
Root cause
Nonconforming Material/Component
Terminated
2020-01-15
Distribution
US Nationwide Distribution
Recall number
Z-2747-2018
Recall status
Terminated
CDRH record ID
165793
FDA record created
2018-08-11
Recall event number
80586

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗