- Recall classification
- Class II
Reason stated in the source
Potential hardware issue which may cause thermal effects and possible damage to system parts.
Source facts
- Action
- On June 22, 2018, Siemens Healthineers issued URGENT MEDICAL DEVICE CORRECTION notices to customers via courier service. In the event the problem occurs, the system can be operated normally, however, in the case of a leakage current, X-ray will not be possible. This may result in a situation where it is necessary to cancel or restart clinical treatment or to continue treatment on an alternative system. Please follow the standard emergency procedures you have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. Siemens will correct the error by updating all affected generators with a RCD (Residual Current Device) to detect leakage currents via Update Instruction AX060/16/S. Additionally, Siemens recommends installing a RCD (Residual Current Device) in the electrical system of the hospital as an appropriate backup system and general electrical protection. You may continue to use the Artis Q and Artis Pheno systems while awaiting implementation of the corrective action. Siemens Healthineers will contact customers to arrange a date to perform this corrective action. Please feel free to contact the service organization for an earlier appointment at 1-800-888-7436.
- Initiated
- 2018-06-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-01-15
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2748-2018
- Recall status
- Terminated
- CDRH record ID
- 165794
- FDA record created
- 2018-08-11
- Recall event number
- 80586