Z-2750-2020: CARESCAPE ONE, Physiological Patient Monitor - Product Usage: is both a multi-parameter physiological pat

Recall record Z-2750-2020 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2750-2020
Company in FDA record
GE Healthcare, LLC
FDA classification date
Aug 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

CARESCAPE ONE may not provide visual and audible alarms for Ventricular Fibrillation (V Fib), if V Fib occurs at the time CARESCAPE ONE is docked to a CARESCAPE B450/B650/B850 host monitor.

Source facts

Action
You can continue to use your CARESCAPE ONE to monitor patients. Follow the instructions below each time the CARESCAPE ONE is docked to a CARESCAPE B450/B650/B850 host monitor. 1. As per the User Manual, closely observe and monitor the patients clinical condition and ECG waveforms on the CARESCAPE B450/B650/B850 host monitor for V Fib during the Learning phase and for several minutes after the Learning phase to allow the algorithm to reach optimal detection performance. 2. As per the User Manual, if the Learning message on the host monitor does not clear after 30 seconds, select the Relearn QRS option from the ECG menu on the host monitor following these steps. Select the HR parameter window. Select Advanced > Page 1. Select Relearn QRS. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. After the CARESCAPE ONE Monitor has been updated, discontinue usage of any previous versions of CARESCAPE ONE software. Destroy any software media containing previous versions of CARESCAPE ONE software, including reimage, upgrade kits, and FRUs. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2020-06-22
Root cause
Software Design Change
Terminated
2024-08-19
Distribution
Worldwide distribution - US Nationwide distribution including in the states of MT, NC, TX and WI. The countries of Australia, Austria, Bahrain,Belgium, Bolivia, Brazil ,Canada, Chile, Check Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Malta, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Russia, Saudi Arabia, Singapore, Saint Martin, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Arab Emirates, United Kingdom.
Recall number
Z-2750-2020
Recall status
Terminated
CDRH record ID
182369
FDA classified
2020-08-04
Recall event number
85979

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗