Z-2751-2018: Monitor with one Blood Parameter Module and one HSat Probe; 500AHCT Monitor with two Blood Parameter modu

Recall record Z-2751-2018 for Terumo Cardiovascular Systems Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2751-2018
Company in FDA record
Terumo Cardiovascular Systems Corporation
FDA record creation date
Aug 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.

Source facts

Action
On June 26, 2018, the firm began notifying customers of the market withdrawal via telephone calls. The call alerted customers of the issue and scheduled an expedited return of affected CDI System 500 units to the firm's service depot so that Service Technicians may replace the affected H/SAT with a replacement H/SAT device. Customers then received a copy of the completed phone script by email along with a customer response form. Customers were asked to review the phone script, and complete and return the response form. Customers were also asked to assure that all users were notified of the issue, and to prominently display the notice where all users may access it. To contact Terumo CVS with any questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 (Monday Friday, 8 a.m. 6 p.m. ET) Recall Fax: 1.734.741.6149
Initiated
2018-06-26
Root cause
Vendor change control
Terminated
2020-04-17
Distribution
Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MS, NC, NM, NY, OH, SC, TN, TX, WA, and WV. Foreign distribution to Middle East, Latin America, Europe, Japan, Canada, and Mexico.
Recall number
Z-2751-2018
Recall status
Terminated
CDRH record ID
165792
FDA record created
2018-08-11
Recall event number
80530

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗