Open original FDA record ↗Recall record
- FDA identifier
- Z-2753-2011
- Company in FDA record
- Kapp Surgical Instrument Inc
- FDA classification date
- Jul 7, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During an FDA Inspection on 3/21/11 - 3/22/11 at Kapp Surgical an uncontrolled instruction for use was discovered which contained sterilization parameters which differed from the instructions contained in the device master record. Kapp Surgical decided to conduct the recall because we cannot be sure that incorrect instructions were not shipped.
Source facts
- Action
- Kapp Surgical Instrument, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 24, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory and quarantine any product affected by this recall. The firm will make arrangements for returning the products. Replacements will be made by Kapp Surgical. Questions about this recall are directed to the QA Manager at 800-282-5277.
- Initiated
- 2011-05-24
- Root cause
- Labeling mix-ups
- Terminated
- 2013-02-19
- Distribution
- Product was distributed to the following states: OH, PA & SC. Product was also shipped to Saudia Arabia.
- Recall number
- Z-2753-2011
- Recall status
- Terminated
- CDRH record ID
- 100856
- FDA classified
- 2011-07-07
- Recall event number
- 59024