Skip to content

Public FDA record

Z-2753-2011: Bariatric Target Ring (Nitinol) (a thin, circular nitinol ring) Model # K-9536-N, Lot #08-001, Lot #08-00

Recall record Z-2753-2011 for Kapp Surgical Instrument Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2753-2011
Company in FDA record
Kapp Surgical Instrument Inc
FDA classification date
Jul 7, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During an FDA Inspection on 3/21/11 - 3/22/11 at Kapp Surgical an uncontrolled instruction for use was discovered which contained sterilization parameters which differed from the instructions contained in the device master record. Kapp Surgical decided to conduct the recall because we cannot be sure that incorrect instructions were not shipped.

Source facts

Action
Kapp Surgical Instrument, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 24, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory and quarantine any product affected by this recall. The firm will make arrangements for returning the products. Replacements will be made by Kapp Surgical. Questions about this recall are directed to the QA Manager at 800-282-5277.
Initiated
2011-05-24
Root cause
Labeling mix-ups
Terminated
2013-02-19
Distribution
Product was distributed to the following states: OH, PA & SC. Product was also shipped to Saudia Arabia.
Recall number
Z-2753-2011
Recall status
Terminated
CDRH record ID
100856
FDA classified
2011-07-07
Recall event number
59024

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.