Skip to content

Public FDA record

Z-2753-2019: Accu-Chek Connect Diabetes Management App

Recall record Z-2753-2019 for Roche Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2753-2019
Company in FDA record
Roche Diabetes Care, Inc.
FDA record creation date
Sep 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Users with Android OS 8.0 and above may be unable to transfer values obtained on their meter to the App.

Source facts

Action
On 5/2/19, the firm notified affiliates of the issue via a Quality Notice. On 5/16/19, the firm issued an updated Quality Notice stating that a new version of the software, 2.1.7.2 would soon become available which would resolve the data transfer issues. Call centers were also provided with a workaround for customers who called in reporting an issue with transferring results. On 5/24/19, version 2.1.7.2 of the Connect App was released for Android on the Google Play store. This version of the app corrects the data transfer issue. At this time, users were notified via the "What's New" section of Google Play. These users were informed that the new version addresses issues with wireless data transfers from blood glucose meters. Customers with questions regarding this event should call the ACCU-CHEK Customer Care Software support line at 1-800-628-3346.
Initiated
2019-05-03
Root cause
Software in the Use Environment
Terminated
2020-10-30
Distribution
Nationwide and worldwide usage of the Connect App.
Recall number
Z-2753-2019
Recall status
Terminated
CDRH record ID
175749
FDA record created
2019-09-26
Recall event number
83566

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗